Peer-reviewed veterinary case report
Evaluation of the agreement between a point-of-care lactate meter and a handheld laboratory analyzer in cats treated in emergency practice.
- Journal:
- Journal of veterinary emergency and critical care (San Antonio, Tex. : 2001)
- Year:
- 2020
- Authors:
- Acierno, Mark J et al.
- Affiliation:
- Department of Medicine
- Species:
- cat
Abstract
OBJECTIVE: The purpose of this study was to determine if there was agreement between a new point-of-care (POC) lactate analyzer and a handheld laboratory analyzer when measuring blood lactate concentration in cats. DESIGN: Prospective observational study. SETTING: University veterinary teaching hospital. ANIMALS: Fifty-four cats that presented to an emergency service. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Lactate concentrations as measured by the handheld laboratory analyzer ranged from 0.3 to 15.4 mmol/L. Agreement analysis of the handheld laboratory analyzer and the POC lactate meter demonstrated a bias, -0.06 and limits of agreement ranging from -0.87 to 0.99 mmol/L. Regression analysis demonstrated a coefficient of determination (R) of 0.98. CONCLUSION: Results of the present study indicate that the POC lactate meter provided results that are in agreement with a handheld laboratory analyzer when measuring lactate in clinically ill patients.
Find similar cases for your pet
PetCaseFinder finds other peer-reviewed reports of pets with the same symptoms, plus a plain-English summary of what was tried across them.
Search related cases →Original publication: https://pubmed.ncbi.nlm.nih.gov/31845516/