Agreement between direct and indirect blood pressure measurements obtained from anesthetized Hispaniolan Amazon parrots.

Journal:
Journal of the American Veterinary Medical Association
Year:
2008
Authors:
Acierno, Mark J et al.
Affiliation:
Department of Veterinary Clinical Science · United States
Species:
bird

Abstract

OBJECTIVE: To determine the level of agreement between direct and indirect blood pressure measurements obtained from healthy Hispaniolan Amazon parrots (Amazona ventralis) anesthetized with isoflurane. DESIGN: Validation study. ANIMALS: 16 healthy adult Hispaniolan Amazon parrots. PROCEDURES: Parrots were anesthetized, and a 26-gauge, 19-mm catheter was placed percutaneously in the superficial ulnar artery for direct measurement of systolic, mean, and diastolic arterial pressures. Indirect blood pressure measurements were obtained with a Doppler ultrasonic flow detector and an oscillometric unit. The Bland-Altman method was used to compare direct and indirect blood pressure values. RESULTS: There was substantial disagreement between direct systolic arterial blood pressure and indirect blood pressure measurements obtained with the Doppler detector from the wing (bias, 24 mm Hg; limits of agreement, -37 to 85 mm Hg) and from the leg (bias, 14 mm Hg; limits of agreement, -14 to 42 mm Hg). Attempts to obtain indirect blood pressure measurements with the oscillometric unit were unsuccessful. CONCLUSIONS AND CLINICAL RELEVANCE: Results suggested that there was substantial disagreement between indirect blood pressure measurements obtained with a Doppler ultrasonic flow detector in anesthetized Hispaniolan Amazon parrots and directly measured systolic arterial blood pressure.

Find similar cases for your pet

PetCaseFinder finds other peer-reviewed reports of pets with the same symptoms, plus a plain-English summary of what was tried across them.

Search related cases →

Original publication: https://pubmed.ncbi.nlm.nih.gov/19014291/